• How To Operate Rotary Evaporator?

    Rotary evaporator
    also known as rotary evaporator, is a common laboratory distillation extraction equipment, widely used in the concentration, drying, refining, separation, crystallization and other industries.


  • Application method of rotary evaporator:


  • First Step:

    1.Adjustment of height: manual lifting, hand wheel above the motor column, turn upward and reverse downward. Electric lift, hand touch up key main engine up, hand touch down key main engine down.


  • Second Step:

    2.Two outer joints of the condenser are used for cooling water, one end is used for water intake and the other end is used for water discharge,and the lower the temperature of the condensate water is,the effect is.


  • Third Step:

    3.Before starting the machine, turn the speed control knob to the minimum left, press the power switch indicator light, and then turn it to the right slowly to the required speed.Generally,medium speed,low speed and high viscosity solutions are used for large vaporator.


  • Four step:

    4.When using, should decompress first, drive motor to turn distillation flask again, when ending, because be out of power first motive,again open atmosphere,prevent distillation flask to fall off in turning.

  • Above is the operation procedure of rotary evaporator. If you want to know more about and experience the real object, please contact: [email protected]
     now.

     